Fig. 9  SAE division 33 single-tooth fatigue (STF) test fixture 
Test Procedure. Special concern must be taken with regard to safety in conducting the STF test (in addition to 
general lab safety procedures). Setting up the test requires moving the hydraulic load ram while lining up the 
fixture, bypassing safety features, such as a light curtain to stop the machine, and so on, and considerable 
caution needs to be exercised during setup. Periodic calibration of the test fixture is important because there are 
many closely fitted parts that can wear and change the bending stress in specimen gears. A calibration gear is 
made from a standard specimen with strain gages fit at key points in the root fillets. This calibration gear is 
installed in the fixture and loaded to set loads periodically to ensure consistent loading. If specimens for a 
specific test program differ from the standard design established for the fixture, one of these specimens should 
also be fitted with strain gages, and calibrations should be conducted with both gears. Typical calibration 
intervals are from every six to every thirty tests, or any time the fixture is cleaned and repaired. 
Other set-up items that can affect test results are mounting of the fixture on the universal test stand, tuning of 
the test-stand controller, and test frequency. If the SAE type fixture is used, it is important to ensure that the 
fixture floats freely on an oil film before starting each test. Other fixture designs can be bolted to the machine 
base, provided proper alignment between the load ram and loading point on the load arm is maintained. The 
feedback control should be tuned to suit test fixture and specimen compliance. Some systems do this 
automatically each time the machine is turned on; with others, this tuning may have to be done manually. At a 
minimum, tuning should be verified as often as fixture calibration. Some systems provide a feature to 
automatically adjust tuning as compliance changes. This feature should be turned off when running STF tests—
a significant change in compliance is proof of failure. The maximum frequency for STF testing is limited by the 
capacity of the system to maintain a satisfactory load waveform. An inability to maintain peaks on the unload 
side of the wave at the specified value is a sign of testing at too high a frequency. All of the tests for a given 
project should be conducted at the same frequency, which is usually set at slightly less than the maximum for 
the highest anticipated load. 
The other procedural item that can affect test results is the method used to detect failure and stop the test (and 
stop counting cycles). All systems can be set up to stop if the load goes out of the specified range. This 
deviation will happen when compliance changes, usually as a result of a crack in the root fillet. Many systems 
incorporate a linear variable differential transformer and can be set up to stop if the load ram moves beyond the 
specified range (also an indication of tooth failure). A limit switch can be set to trip when the load ram moves 
too far. Other failure detection devices, such as a crack wire bonded near the root fillet so that it will break 
when the tooth cracks, or an ultrasonic system to detect cracking, can be used to stop the test. At least two of 
these methods should be employed to ensure that the machine stops cycling load when the tooth breaks. If a 
computer is used to control loading during tests, one of the failure detection systems should be connected 
directly to the universal test-stand controller to guarantee that the machine stops if the computer crashes. 
Specimen Results. Table 2 summarizes results from a typical set of STF tests. Testing was conducted in three 
phases. Initial searching tests were conducted to establish loads that would result in failure in reasonable time. 
A “modified staircase sequence” of tests was conducted to develop data at a series of loads representing zero to 
100% failure. Further tests were conducted to fill in the stress cycles relationship. Searching tests are started at 
a high load to ensure starting with a failure, then stepped down until the tooth survives the specified number of 
cycles (here, 5 million cycles has been selected as a run-out limit). The modified staircase sequence is 
conducted by testing three specimen gears in sequence. If the tested tooth breaks before the specified limit, the 
next test is conducted one load step lower. If it does not break by the specified limit, the next test is conducted 
one load step higher. After the modified staircase sequence is completed, additional tests are conducted to 
ensure that all the specimen gears are tested at the lowest load. More tests are conducted to develop enough data 
for Weibull analysis at two loads resulting in 100% failure. 
Table 2   Results from a typical set of gear single-tooth fatigue tests showing overall testing sequence, 
including modified staircase (“up and down”) tests 
Legend: X, failure; O, runout. Specimen serial numbers 2, 4, and 6. Specimens cut from bar stock, hobbed 
roots. R, loading; R = 0.1. Frequency, 25 Hz