68
used for STEM, either a cold fi eld emission gun (CFEG) or a Schottky
thermally assisted fi eld emission gun. In the case of a CFEG, the source
size is typically around 5 nm, so the probe-forming optics must be
capable of demagnifying its image of the order of 100 times if an atomic
sized probe is to be achieved. In a Schottky gun the demagnifi cation
must be even greater.
The size of the image of the source is not the only probe size defi ning
factor. Electron lenses suffer from inherent aberrations, in particular
spherical and chromatic aberrations. The aberrations of the objective
lens generally have greatest effect, and limit the width of the beam
that may pass through the objective lens and still contribute to a small
probe. Aberrated beams will not be focused at the correct probe posi-
tion, and will lead to large diffuse illumination thereby destroying
the spatial resolution. To prevent the higher angle aberrated beams
from illuminating the sample, an objective aperture is used, and is
typically a few tens of microns in diameter. The existence of an
objective aperture in the column has two major implications: (1) As
with any apertured optical system, there will be a diffraction limit
to the smallest probe that can be formed, and this diffraction limit
may well be larger than the source image. (2) The current in the
probe will be limited by the amount of current that can pass through
the aperture, and much current will be lost as it is blocked by the
aperture.
Because the STEM resembles the more commonly found SEM in
many ways, several of the detectors that can be used are common to
both instruments, such as the secondary electron (SE) detector and the
energy-dispersive X-ray (EDX) spectrometer. The highest spatial reso-
lution in STEM is obtained by using the transmitted electrons, however.
Typical imaging detectors used are the bright-fi eld (BF) detector and
the annular dark-fi eld (ADF) detector. Both these detectors sum the
electron intensity over some region of the far fi eld beyond the sample,
and the result is displayed as a function of probe position to generate
an image. The BF detector usually collects over a disc of scattering
angles centered on the optic axis of the microscope, whereas the ADF
detector collects over an annulus at higher angle where only scattered
electrons are detected. The ADF imaging mode is important and unique
to STEM in that it provides incoherent images of materials and has a
strong sensitivity to atomic number allowing different elements to
show up with different intensities in the image.
Two further detectors are often used with the STEM probe stationary
over a particular spot: (1) A Ronchigram camera can detect the inten-
sity as a function of position in the far fi eld, and shows a mixture of
real-space and reciprocal-space information. It is mainly used for
microscope diagnostics and alignment rather than for investigation of
the sample. (2) A spectrometer can be used to disperse the transmitted
electrons as a function of energy to form an electron energy-loss (EEL)
spectrum. The EEL spectrum carries information about the composi-
tion of the material being illuminated by the probe, and even can show
changes in local electron structure through, for example, bonding
changes.
Peter D. Nellist