January 9, 2009 10:21 World Scientific Book - 9.75in x 6.5in ws-bo ok975x65˙n˙2nd˙Ed
600 Principles of Radiation Interaction in Matter and Detection
are within the range of ±5 ADC counts.
The reference pulse position is used to correct the gain of the preamplifier at
each electric field value. It is seen from Fig. 8.5(b) that the reference peak positions
are varying during the measurements within the range of ±2 ADC around the
average value. This conclusion was checked to be valid over the whole ADC scale.
The calibration of the ADC scale in terms of collected charge units (fC) is
obtained from the linear fit of the calibration pulse positions. The fitted parameters
are averaged over a set of measurements, and the deviations from this averaged
calibration are presented in Fig. 8.5(c). Based on this calibration method, the un-
certainty on the collected charge value is expected to be at the level of 0.01 fC.
The detailed analysis of the systematic errors was performed for all irradiation
runs. Some significant variations of the systematics can be found due to cryostat
activation. The activation of the cryostat is at its lowest after a long stopping
perio d (3–4 months) of the reactor, due to maintenance. During normal period of
operation, frequent irradiation every three weeks increases the cryostat activation
and thus the systematic errors. In addition, this activation maintains some level of
impurities in liquid argon before irradiation. It is important to know the evolution
with time of this impurity level. A special test was dedicated to the monitoring over
a long period of time of the evolution of the impurity level of liquid argon purposely
contaminated with oxygen in the cryostat. About 6 cm
3
of oxygen gas were injected
into the warm argon vessel of the cryostat which was pumped out beforehand for this
test. A maximal value of about 11 ppm of oxygen concentration might be expected
in this case for the whole oxygen probe liquefied in the cryostat, and about 7 ppm
for the oxygen probe mixed uniformly in the liquid and gaseous states of argon
in cryostat and receiver. The oxygen concentration values were determined using
Eqs. (8.37–8.39). The results are presented in Fig. 8.6.
The liquid argon purity improves with time after the initial liquefaction of the
oxygen probe due to the mixing of the oxygen polluted liquid argon in the cryostat
and the argon gas in the receiver. The initial value of the impurity concentration
is close to expectation (11 ppm). The “cleaning” could continue until the oxygen
concentration in the argon system becomes even. To speed up the process, the
oxygen polluted liquid argon was evaporated from the cryostat back into the receiver
and liquefied once more. The liquid argon was kept in the cryostat after this full
mixing until the end of the test (6 days) and the liquid argon purity remained stable
within ±1 ppm, close to the expected level of ∼ 7 ppm. It was found that variations
of the monitor signal at the end of the test could be caused by the instability of the
high voltage power supply.
8.3.3 Examples of Pollution Tests Using the α-Cell
A very first test consisted of a so-called “empty” run. The cryostat filled with
liquid argon (with no material for test inside, but everything else being identical to