7.2 The Scanning Tunneling Microscope (STM)
r0.5 volt to r10.0 volts, with the ability to set asymmetrical limits being
highly desirable), or in z span (from about –2 nm to +5 nm, where negative
distance moves the tip closer to the substrate than at the set-point. The most
commonly used auxiliary input is the output of a lock-in amplifier that detects
a modulated signal in the tip current.
2) It must be possible to add modulation (typically sine or square wave) to
the bias or tip position voltage. An old version of the Digital Instruments
(now Veeco) software, version 3.2x, had built in square wave modulation and
the software could display directly dI/dV, determined from the resulting
modulation in the current. This system worked very well, but was dropped
from, or incorrectly implemented in, later versions of the software. To our
knowledge, none of today’s manufacturers offer such complete “built-in”
spectroscopic capability.
3) The control system must be able to shut off the feedback loop during
data acquisition, when desired.
4) The drift (both x, y and z) must be low. The extent of acceptable x, y
drift is determined by the amount of spatial resolution desired for the
spectroscopic data. If one is acquiring spatially averaged molecular spectra
from a dense monolayer with a large curvature tip, drifts of the order of
atoms through their vibrational signature in STM-based IETS, drifts of less
than 0.05 nm/min are required. The amount of acceptable drift in the z
direction is determined by the intensity of the spectral feature to be studied
relative to the background tunneling current in the absence of that feature.
Since this background increases (approximately) as I and I = CV exp(–
Az)), a small change in z, dz, results in an increase in relative current (dI/I)
of –A)dz. Or, using z in nm and ) in volts, the background change in
relative current with a small z drift of dz is given by dI/I ~ –0.1 dz. For true
vibrational IETS (with no resonance enhancements), dI/I for the spectral
transition, dI/I ~ 0.002. Thus, less than 1 × 10
–2
nm of z drift is allowed during
the time required to scan a particular spectral band. For STM-OMTS, on the
other hand, dI/I ~ 0.1, and 100 times as much z drift is allowed. Moreover,
because the OMTS signal is so much stronger than that of IETS, equal signal-
to-noise can be obtained about (0.1/0.002)
1/2
~ 7 times faster. Overall, STM-
OMTS is expected to be about 1000 times less z drift-sensitive than STM-
IETS of non-resonance enhanced transitions.
5) Sample and tip geometry and shielding: Because of the drift constraints
discussed above, one needs to take spectra as fast as the electronic bandwidth
allows. With the feedback loop turned off, the limiting term is the capacitance
in the tip-sample assembly and the wires leading to the preamplifier circuit.
Thus, the preamplifier needs to be close to the tip, and the tip and sample need
to be electrically isolated and as small as possible.
6) An essential requirement for STM-IETS is that the working parts of the
STM be at or below 10 K. Otherwise, the thermal line-width destroys the
information inherent in identifying vibrational peak positions. It is not enough
1 nm/s can be tolerated. On the other hand, if one is trying to locate particular
311