
236 6 Detectors for Photon Counting
The XP2020UR (shown in the middle) has a wider SER and a slower leading
edge. Although the operating voltage is far below the permissible maximum, the
average pulse amplitude is more than 100 mV. At first glance the high pulse am-
plitude may be considered a benefit. In practice, the high amplitude in conjunction
with the broad SER is rather a drawback, because it leads to a high average output
current at a given count rate. The high output current distorts the dynode voltage
distribution at high count rates, which in turn causes the TCSPC instrument re-
sponse to be count-rate-dependent; see Fig. 7.33, page 296.
The SER of an R931 side-window PMT is shown in Fig. 6.21, right. Side-
window PMTs were not originally designed for fast detection, and a large amount
of ringing is present in the SER. However, the somewhat ugly SER pulse shape
has no negative effect on TCSPC applications. Although the ringing may trigger
the CFD several times, this happens within the dead time of the TCSPC module
and is therefore not recorded. The situation is different for fast multichannel
scalers with a dead time of the order of 1 ns. If the time-channel width is of the
order of 1 ns, the ringing shows up in the IRF. Even if the time-channels are wider
than 1 ns, multiple counting of the larger pulses makes it impossible to obtain a
clear counting plateau by adjusting the discriminator threshold.
6.3.2 Transit Time Spread
The transit time spread is measured by illuminating the PMT cathode with short
light pulses and recording the pulse density versus time in a normal TCSPC setup.
The setup is the same as for recording of the instrument response function (IRF) of
a TCSPC system, and the same precautions should be taken to avoid broadening
of the measured response (see Sect. 5.1.6, page 75). Sufficiently short pulses have
to be used, and any optical elements that can cause pulse dispersion must be re-
moved. Monochromators, optical fibres, diffusers or scattering solutions must be
strictly avoided. The light of the test laser should be sent directly to the cathode of
the PMT through a set of neutral density filters. The stop pulse for the TCSPC
module should be delayed so that the same laser pulse that caused a photon also
stops the TAC. The count rate must be kept low enough to avoid pile-up. The
CFD threshold should be set to detect at least 20% of the detector pulses to avoid
distortion by multiphoton events (see Fig. 7.62, page 320).
A frequent source of failure in TTS measurements is improper shielding of the
detector. RF noise pickup from the environment or line frequency pickup via
ground loops can severely impair the timing accuracy of the TCSPC electronics. It
is obvious that the correct TTS cannot be measured under such conditions.
The TTS of conventional PMTs and miniature PMTs with metal channel dyn-
odes can be measured with satisfactory accuracy using picosecond diode lasers.
These lasers deliver pulses as short as 30 to 50 ps FWHM. However, the pulses
may have a tail or a shoulder, especially at higher power. The diode driving condi-
tions for clean pulse shape with minimum tail are usually not the same as for
shortest FWHM. The TTS of MCP PMTs can be reasonably measured only by a
Ti:Sapphire laser or a similar femtosecond or picosecond laser system.