
Experimental techniques  323 
Hence,  a  continuous  scan  produces  data  nearly  identical  to  those 
collected by means of a step scan, i.e. powder 
diffraction 
data are saved in 
the format shown in Figure 
3.43. 
The only difference is that the intensity is 
not given for a fixed detector position, but for a median Bragg angle in the 
scanned interval. To minimize the  introduction of  a small, but systematic 
error, an intensity measurement during continuous scanning always begins 
from 20 
= 
20mdia, 
- 
A2012, where 29,dia,,  is the median Bragg angle saved in 
the data file. For example, the diffracted intensity at the Bragg angle 20 
= 
10" with a sampling interval A20 
= 
0.02" is the result of accumulating an 
x-ray photon count during continuous scanning from 
9.99 
to 10.01" of 20. 
The  two  most  important parameters  in  a  continuous scan,  which  are 
defined by  the  user,  are  the  sampling interval  (step), s, and  the  angular 
velocity (scan rate), r. The sampling step is equivalent to the step size in the 
step scan mode. Everything said about the size of the step in the previous 
section, therefore, applies to the sampling step during the continuous scan. 
The two parameters, i.e. counting time, t, in the step scan and the scan rate, 
r, in the continuous scan are related to one another as follows 
In 
Eq. 3.7, t is in seconds, 
s 
is in degrees, and r is in degreeslmin. Thus, a 
continuous scan with the rate r 
= 
0.1 deglmin and with the sampling step 
s 
= 
0.02" is equivalent to a step scan with the same step and counting time 
12 slstep. When the sampling step is reduced at a constant scan rate, this is 
equivalent to the proportional reduction of counting time and vice versa. 
In 
modern diffractometers both scanning modes result in nearly identical 
quality of experimental data. 
A 
step scan is usually considered as the one 
with  less  significant  positioning  errors,  which  could  be  important  in 
experiments where the  maximum  lattice parameter precision  is  essential. 
Continuous scans are used  most  often  for  fast experiments, whereas step 
scans are usually employed in overnight or weekend experiments. 
Predictably, when  counting times are short (t 
<< 
1 
s) step scans take 
longer to complete (the required time may be easily doubled when compared 
with the identical quality continuous scans). This occurs because no intensity 
is measured when the goniometer arms move to the next position (compare 
the flow chart from Figure 3.42 with that from Figure 3.45). The difference 
in  the  time  of  the  experiment becomes  negligible  during  overnight  or 
weekend experiments. 
Many commercial powder  diffractometers have physical  limits on the 
lowest scan rate. For example in the Scintag 
XDS2000 system, the scan rate 
cannot be  lowered below 0.1 deglmin. This scan rate is equivalent to the