
248 Measurement and Data Analysis for Engineering and Science
by down-scale output values that are higher than their up-scale counterparts.
The best-fit curve of the data is indicated by a solid line. Both the hysteresis
and linearity errors are assessed with respect to the best-fit curve.
Sensitivity, repeatability, zero-shift, stability, and thermal-drift errors
are ascertained by performing a series of calibrations and then determining
each particular error by comparisons between the calibrations. The results
of a series of calibrations are shown in the right graph of Figure 7.3. The
solid curve represents the best-fit of the data from all the calibrations. The
dotted curves indicate the limits within which a calibration is repeatable
with 95 % confidence. The repeatability error is determined from the differ-
ence between either dotted curve and the best-fit curve. The dash-dot curves
identify the calibration curves that have the maximum and minimum slopes.
The sensitivity error is assessed in terms of the greatest difference between
minimum or maximum sensitivity curve and the best-fit curve. The dashed
curves denote shifts that can occur in the calibration because of zero-shift,
stability, and thermal-drift errors. Each error can have a different value and
is determined from the calibration curve having the greatest difference with
calibration data that occurs with each effect, with respect to the best-fit
curve.
The following two examples illustrate the effects of instrument errors on
measurement uncertainty.
Example Problem 7.5
Statement: A pressure transducer is connected to a digital panel meter. The panel
meter converts the pressure transducer’s output in volts back to pressure in psi. The
manufacturer provides the following information about the panel meter:
Resolution: 0.1 psi
Repeatability: 0.1 psi
Linearity: within 0.1 % of reading
Drift: less than 0.1 psi/6 months within the 32
◦
F to 90
◦
F range
The only information given about the pressure transducer is that it has “an accuracy
of within 0.5 % of its reading”.
Estimate the combined standard uncertainty in a measured pressure at a nominal
value of 100 psi at 70
◦
F. Assume that the transducer’s response is linear with an
output of 1 V for every psi of input.
Solution: The uncertainty in the measured pressure, (u
d
)
mp
, is the combination of
the uncertainties of the transducer, (u
d
)
t
, and the panel meter, (u
d
)
pm
. This can be
expressed as
(u
d
)
mp
=
q
[(u
d
)
t
]
2
+ [(u
d
)
pm
]
2
.
For the transducer,
(u
d
)
t
=
q
u
2
I
t
+ u
2
o
t
= u
I
t
= 0.005 × 100 psi = 0.50 psi.
For the panel meter,
(u
d
)
pm
=
q
u
2
I
pm
+ u
2
o
pm
.