
Magnetic Properties 10.3 Magnetic Characterization in a Pulsed Field Magnetometer (PFM) 573
that of Cu. Therefore the error in magnetization meas-
urements due to eddy currents is rather small. These
considerations have led to the development of eddy-
current correction for hysteresis loops measured in
pulsed fields, which is called the f/2 f method. In this
case one measures the loop with two different pulse
durations and calculates the corrected loop point by
point, applying an extrapolation procedure [10.37, 38].
It was shown by finite-element calculations that for
eddy-current errors that were not too large (less than
20%) the corrected loop agrees with the true loop within
2%. This means that the effect of eddy currents is under-
stood and can be corrected in most cases [10.39].
b) Magnetic viscosity effects. When the hysteresis loop
of hard magnetic materials is measured in transient
fields the so-called magnetic viscosity causes a differ-
ence between the measured loop and the true loop.
The magnetic viscosity is also observed in nonconduct-
ing materials (e.g., ferrites), therefore it is not due to
eddy currents. It has to be mentioned that the time
constant of the exponential decay of of eddy currents
(in metallic samples) is of the order of typical val-
ues of microseconds, whereas that of the logarithmic
decay of the viscosity lies between milliseconds and
seconds. Additionally eddy currents depend on the ge-
ometry of the sample whereas this is not the case for the
viscosity.
10.3.3 Calibration [10.1]
Field Calibration
The field is calibrated using a small pickup coil whose
effective winding area is known from an NMR cali-
bration. The induced voltage u(t) is then fitted using
(10.53), in order to determine the field calibration
factor k, the damping factor and the pulse duration (in-
cluding the effect of the damping)
H = H
0
exp(−at) sin ωt ,
u
i
(t) =−NA
dB
dt
=−NAμ
0
H
0
d
dt
exp(−at) sin ωt
. (10.53)
Table 10.3 Summary of calibration results
Sample Shape μ
0
H
max
(T) μ
0
M μ
0
M
literature
Error μ
0
M (T)
t = 57 ms (T) (T) (%) t = 40 ms
Fe
3
O
4
Sphere 2r = 5.5mm 1.5 0.5787 ±0.001 0.569 [10] +1.6 0.5782
Ni Cylinder D = 4;h =8mm 1.5 0.6259 ±0.0008 0.610 [11] +2.6 0.6322
Fe Cylinder D = 4;h =8mm 4.5 2.1525 ±0.0051 2.138 [12] +1.4 2.1826
The damping factor a determined in this way can be
compared with the value given by a PFM circuit us-
ing a ≈ R/2L (where R is the resistivity of the pulse
magnet, and L is the inductivity of the pulse magnet).
The calibration factor k is also determined as
a function of the gain (integrator gain and the gain of
preamplifier). The calibration factor k gives a relation
between the induced voltage and the field at the search
coil. At the same time the integrated voltage (using dif-
ferent gain factors) of the H-measuring coil (which is
located at the pickup rod) on the magnetometer system
is recorded.
It was shown that the calibration factor k was also
determined as a function of the gain using an analogue
integrator [10.1]. The scattering of the k factor was be-
low ±1%. This indicates that the linearity of the gain
is better than 1%. Using such a procedure, an absolute
field calibration of better than 1% is achieved, including
the time constants (gain) of the integrator.
Magnetization Calibration
The magnetization is calibrated using well-known ma-
terials such as Fe and Ni (in which the eddy-current
error causes an uncertainty) or preferably a nonconduct-
ingsamplesuchasFe
3
O
4
or a soft magnetic ferrite,
such as 3C30 (Philips). All calibration measurements
are performed at room temperature. The results of
the magnetization calibration measurements are sum-
marized in Table 10.3. To check the reproducibility
all measurements were repeated 10 times to give an
average value M. Additionally measurements using
a shorter pulse duration (40 ms) were performed, which
were generally in good agreement with that of the long
pulse. For the metallic samples an error of 1–2% due to
the eddy currents occurs.
The mean value of the deviations D
mv
= 1.6% is
higher than the true values. It should be mentioned that
no significant differences in the measured magnetiza-
tion values were obtained, when different pulse duration
were employed.
The mean value of the deviations D
mv
=1.6% has
a standard deviation of 0.95%. The standard deviations
concerning the reproducibility gave a mean value of
Part C 10.3