272 Getting started
r
place the carrier in the fixing bath and fix for 8–10 min, again depending on the type and
age of the fixing solution; use frequent agitation;
r
rinse in room-temperature running water for 20–30 min, apply a soapy solution to mini-
mize drying marks (optional), and dry in a warm and dust-free place.
The details of this procedure may vary depending on the type of chemicals and
negatives used. For nearly all micrographs reproduced in this book we used Kodak
Electron Image Film SO-163 with 2:1 diluted D19 developer.
If you plan to digitize your negative, you should use a good flatbed scanner with
a transparency adaptor, and scan the negative at a resolution of 1200 or 2400 dpi.
The grain size in the photographic emulsion is then about equal to the pixel size of
the digitized image. There is little point in using a higher resolution, unless you use
special film with a smaller grain size. Once you have a digital version of the image,
you can use the full power of image analysis programs to enhance the contrast,
scale the brightness linearly or non-linearly, select portions of the image, place a
micron bar on the image, and so on. If you alter the image in any non-standard way
(i.e. adding or removing information, enhancing features, etc.), then you should
explicitly state this in the figure caption (assuming that you are going to use the
figure in a report or paper).
4.5 Microscope calibration
As with any other scientific instrument, a transmission electron microscope must
be calibrated. The image magnification and diffraction camera length are perhaps
the most important calibrations for conventional TEM observations. In addition,
the relative orientation of the image and the corresponding diffraction pattern must
be calibrated. The following two subsections will describe briefly how these cali-
brations can be performed. It is likely that calibrations will have already been
carried out for the microscope(s) you are using, so it may not be necessary to repeat
them yourself. However, you should make sure that you understand the imaging
conditions used for the calibrations, so that you can reproduce them in case you
wish to make accurate measurements.
Calibrations must be carried out not only for conventional micrographs, but also
for digital and analog cameras. In all cases it is important to know the absolute
orientation of the camera/negative with respect to the column. For a negative it is
rather easy to find out how it is oriented inside the column (the emulsion side faces
the electron beam, and there is usually a fiducial marker along the side of the negative
which indicates its orientation in the camera). Calibration measurements should
always be carried out with the negative placed in the same orientation as it was
inside the microscope. Note that it is not sufficient to calibrate magnifications for
electron micrographs alone; if your microscope is equipped with a digital camera,