4.4 A typical microscope session 255
beam moves upon changing this current, then the condensor aperture should be
translated until no further movement occurs for a wide range of beam diameters.
An application which needs a high beam current should use a large-diameter con-
densor aperture, whereas observations which need a high beam coherence should
sacrifice some current by selecting a smaller diameter aperture.
(iii) Illumination stage alignment. This step ensures that the electron trajectory is aligned
with the optical axes of both first and second condensor lenses. The alignment pro-
cedure is usually an iterative procedure whereby the spot size (first condensor lens)
is alternated between two settings, and the beam displacement on the screen is min-
imized by means of the beam and gun shift deflection coils. The beam alignment
is satisfactory when no substantial beam shifts are observ
ed when the spot size is
changed.
†
Figure 4.8(c) shows a typical series of spot sizes for a JEOL 2000EX
microscope.
(iv) Condensor astigmatism correction. The incident electron beam should have a circular
cross-section for all settings of the second condensor lens. Condensor lens astigma-
tism can be recognized readily as an elliptical distortion of the beam, which changes
major axis when the second condensor lens current goes from an under-focus to an
over-focus condition, as shown in Fig. 4.8(d). The condensor stigmator coils must
be used to correct the beam
shape. Condensor astigmatism can also be corrected
in the in-focus condition (smallest beam diameter) by ensuring that the filament tip
produces the sharpest possible image (under-saturated condition).
(v) Filament saturation. When all four previous steps have been carried out properly,
the filament current (in the case of a thermal emission gun) can be increased to the
saturation level, often indicated on the microscope console by the position of a so-
called beam-stop, a metal rod which prevents the user from turning the current up
any further.
(vi) Sample eucentricity. The height of the sample inside the microscope column is an
important variable. The primary tilt axis is a fixed axis with respect to the microscope
column; when the primary tilt is changed, the entire goniometer stage rotates around
this axis. The sample holder
axis, i.e. the center line along the holder rod, does not
necessarily coincide with the primary tilt axis, as illustrated in Fig. 4.9(a). When the
goniometer is tilted, a feature on the sample, indicated by a black circle, will also
rotate around the primary tilt axis, and its projection on the viewing screen will move
side
ways.
‡
The eucentric position is then obtained by adjusting the height of the sam-
ple, the z-control, until there is no further lateral movement during tilting (Fig. 4.9b).
This adjustment could be manual or electronic, depending on the microscope model.
Eucentric alignment is important because it facilitates sample tilting while in diffrac-
tion mode (the region of interest will stay on or near the optical axis). The eucentric
†
Note that the beam diameter will change when changing the spot size; this is due to the fact that the object
plane of the second condensor lens changes with spot size, as discussed in Chapter 3. One must then modify the
second condensor lens current to obtain the same beam diameter on the screen as before.
‡
This is actually useful, because it provides a simple way to identify the orientation of the primary tilt axis on
the viewing screen.