10.2 The microscope as an information channel 599
the function sin χ (q) for the same dimensionless defocus values as Fig. 10.4. The
function oscillates wildly for large values of q, but there is a range of defocus
values D for which the function changes relatively slowly for smaller values of q.
In particular, around D = 1 the function sin χ(q) is close to −1 for a reasonably
large range of q. This range of q is known as a passband, since the value of sin χ(q)
is nearly constant and equal to −1 in this range. The value of cos χ (q) is then nearly
zero. Let us now assume that we can tune the microscope defocus value so that
sin χ(q) =−1 for all values of q; this is obviously another approximation, since
the q-range for which this can be done is finite, not infinite. The aberrated wave
function under this assumption becomes
φ
a
(q) ≈ 1 − σ V
p
(q). (10.21)
In the image plane we find
φ
a
(r) = F
−1
[
φ
a
(q)
]
= 1 − σ V
p
(r). (10.22)
The image intensity is then given by
I (r) =|φ
a
(r)|
2
= 1 − 2σ V
p
(r) + σ
2
V
2
p
(r) ≈ 1 − 2σ V
p
(r), (10.23)
where the quadratic term in σ V
p
is neglected, since we only keep terms linear
in V
p
.
Equation (10.23) can be interpreted as follows: we know from Section 2.6.7
that the interatomic potential V (r) has sharp peaks at the atom positions (see also
Fig. 2.12). The projected potential along a given zone axis direction will also show
large positive peaks at the projected atom positions. Therefore, the intensity I (r)
will be less than unity at the projected atom positions. In other words, if we can find
a microscope setting for which sin χ(q) =−1 over a significant range of q, then
the corresponding image will show dark regions at the projected atom positions.
This is known as dark atom contrast. In the following section, we will explore
how closely we can actually approximate this idealized microscope setting in a real
microscope.
10.2.1.4 The microscope point resolution
The function sin χ(Q) is shown as a grayscale plot in Fig. 10.7. White (black) areas
correspond to positive (negative) values. It is clear that this function oscillates
rapidly for large spatial frequencies Q, as is also obvious from Fig. 10.6; for small
spatial frequencies, the behavior for under-focus and over-focus conditions is dif-
ferent. For under-focus (D > 0) conditions, the first minimum (dark band) appears
for D ≈ 1. At this point the function sin χ(Q) reaches its minimum of −1, which