
10.2 The microscope as an information channel 619
and the floor on which the microscope is mounted. Often, the entire column sits on an
acoustically isolated concrete block. Finally, the microscope operator represents both
a heat source and a source of acoustic vibrations, so ideally the operator should control
the microscope from a remote location. While this is possible in the latest microscope
models, in most situations the operator must be in the same room as the instrument.
Recording images which contain high spatial frequency information therefore requires
some self-discipline from the operator: sounds must be minimized, and preferably the
operator should not move around at all since this will set up air currents and thermal
gradients. For more details on the environmental considerations of high-resolution
work we refer the reader to Chapters 9 and 10 in [Spe88]. A nice example of the
importance of drift and vibration considerations in the design of a microscopy facility
can be found in [Lic01].
(iii) Detector envelope function. The point spread function T (r) (PSF) of the microscope
is a sharply peaked (complex) function with oscillating tails which depend strongly
on the objective lens defocus value. An example PSF for a 400 kV microscope with
C
s
= 1 mm at Scherzer defocus is shown in Fig. 10.17(a). The real and imaginary
parts of both PSF and CTF are shown for a beam divergence angle of θ
c
= 0.8 mrad,
and a defocus spread of 6 nm. Fig. 10.17(b) shows similar
curves for a 200 kV
instrument with identical spherical aberration, but with θ
c
= 0.4 mrad and = 3 nm.
This latter machine has a more coherent electron source which is reflected in a wider
PSF. If the incoherent envelope functions extend to higher spatial frequencies,
then
the corresponding PSF will be wider, and this has consequences for digitally recorded
images.
Consider a CCD camera with M × M pixels. The camera captures an image for a
particular microscope setting, i.e. a particular PSF. The information originating from
a single point in the exit plane of the sample will be spread over an area determined
by the lateral extent of the PSF. The more coherent the source, the larger this area
becomes, i.e. the more delocalized the information becomes. This means that image
pixels which are close to the edge of the CCD camera will detect electrons which
originated from exit plane points that are not part of the acquired image. Figure 10.18
shows schematically that there is a central area on the CCD camera that detects no
electrons from outside the region of interest. The total width of the “incomplete” edge
region is N pixels. The finite size of the detector causes a decrease in the information in
the captured image, and hence the detector itself has an associated attenuation envelope.
Since the width of the PSF is determined by the spatial incoherent envelope function
E
s
(q), it should come as no surprise that the information limit imposed by the detector
depends on the same factors as the spatial incoherence information limit. For a detailed
derivation we refer the interested reader to [dJVD93]. The detector envelope function is
given by
E
D
(q) = exp
+
−2π
2
aC
s
λ
3
q
4
2
N
2
,
, (10.63)