
4.3 LIQUID CALIBRATION FACILITIES 69
Dry Calibration: Dry calibration is not strictly calibration. An orifice meter, if
correctly constructed according to the standards, can be measured, and from the
measurements it should be possible to deduce the flow for a particular pressure
drop.
The same is true of other standard meters such as the venturi, nozzle, and
critical flow venturi nozzle. Other meters are less predictable. No flow test is used,
and so dry
calibration
is something of a misnomer, since we are really talking about
the ability of standards or theory to predict performance. The reader is referred to
the appropriate chapter for the likely predictability of particular flowmeters such
as the electromagnetic and ultrasonic.
4.3 LIQUID CALIBRATION FACILITIES
4.3.1 FLYING START AND STOP
Figure 4.5(a) shows the outline of such a facility, which is most commonly used for
water. This is known as the static weighing method because the water is diverted
into a tank for a fixed period, and the tank can then be measured with the water
static in the tank. In its ideal form, there would be a header tank of large enough size
so that change in level was negligible.
A
header tank, with a weir to retain constant
head, is shown in Figure 4.5(b). The flow from this, by gravity, would pass through
the flowmeter under test and thence to a control valve before discharging into a
diverter valve. This valve would either allow the flow to drop into a weighing tank
or into the sump tank. Water in the weighing tank would be drained into the sump
tank once the weight had been taken. The beauty of this design is that the flow
would be extremely steady due to the header tank and errors would be restricted to
the measurement of the time during which the flow was diverted into the weighing
tank and to the weight
itself.
Figure 4.6 shows a diagram of the diversion process and
the source of timing error. Mattingly (1982) elaborates on the error, and the reader
is referred to the original article for full details.
The essential problem is that it is physically impossible to divert the flow to
and from the weighing tank instantaneously. The time at which the diversion takes
place is, therefore, not at a precise moment, and the changing flow is illustrated in
Figure 4.6. The timer is set to start when volume a equals volume b. This is when
the lost water shown by volume b is just balanced by the gained water in volume
a, which passed before timing began. The same compromise is needed at the end
of the timing period with c equal to d. The setting is often, but not necessarily, the
same at the beginning and end of the diversion (cf. Buttle and Kimpton 1989 for
the design of a diverter).
In order to check the validity of this compromise method, one test is to divert
many times after short periods, comparing the result with one diversion over
a
period
equal to the total of the short runs. This will tend to emphasize the errors in the
diversion, and an adjustment can be made to minimize them. However, there may
be variation in the diversion flow depending on flow rate, and this will require a
more sophisticated compensation method.
It is reasonable to assume that the time measurement can now be several orders
of accuracy higher than the other measured quantities. The errors will therefore