
© 1999 by CRC Press LLC
spring constant and if we set the force to 10
–8
N, then the static deflection of the cantilever is 10 nm. The
typical size of a tunneling force detector is 1-cm length from the tunnel junction to the common
attachment plane. A design with well-compensated thermal expansion coefficients will have a remaining
thermal expansion coefficient of 10
–6
1/K. This means that keeping the force within 10% requires a
thermal stability of the microscope of 0.1 K. In less well compensated design, the allowable temperature
fluctuations might be as low as 0.01 K.
If the temperature stability of the setup is not sufficient, one can either use larger static deflections,
which means larger forces, or a softer cantilever spring, which means degraded frequency response. For
measurements with the smallest possible forces, a careful design of the force sensor with respect to thermal
drift is a prerequisite.
The classical Michelson or Mach–Zehnder interferometers have the worst temperature drift, since their
relevant distances for differential thermal expansions may be more than 10 cm long. The fiber-optic
interferometer is comparable to the tunneling detector in its thermal performance, since the distances
needed to position the end of the fiber are of order 1 cm. Much better is the Nomarski detector for the
cantilever deflection. This detector is only sensitive to a thermally induced rotation of the cantilever
spring. A crude estimate gives relevant distances for the thermal expansion of a few 10 mm. This increases
the allowable temperature variations to more than 1 K.
Equally well suited is the optical lever method. This method is, to first order, only sensitive to the tilt
of the reflecting mirror. For small angles between the incident and the reflected light beam, the change
in distance between the plane defined by the quadrant detector and the light source is negligible. Any
distance change between the light source and the quadrant detector directly affects the output signal.
However, the deflection of the cantilever is amplified by a factor of up to 1000 due to the geometric
amplification. Hence, the optical lever method is, to first order, insensitive to thermal drift.
2.9.2 Classical Setup
Scanning tunneling microscopes are almost exclusively built according to one principle: the tunneling
tip is mounted on some translation stage and scanned past the sample. The reason for this is twofold.
First, most STMs were built for ultrahigh vacuum operation, where it is of paramount importance that
the sample can be exchanged. Since piezos are rather fragile, it was the natural choice to scan the tip.
The second reason is that the tip is lighter than the sample. As one can deduce from the equations
describing cantilevers, reduced masses mean increased resonance frequency and hence increased scanning
speed. By following these rules, AFMs were built with scanned samples. Most AFMs operate at ambient
conditions. The force detection mechanism is bulky; especially after the implementation of the optical
lever (Meyer and Amer, 1988; Alexander et al., 1989), where the detection unit is of the size of several
centimeters, microscope designers preferred to scan the sample instead of the sensor..
Figure 2.23 shows an example of such a microscope. It consists of two parts, the base and force sensor
head. The base houses the scanning piezo, the motor driving the approach screw, and the sample. All
connectors from the control unit to the microscope end here. The base acts as an additional vibration
damper, shielding the sensitive parts from the environment. The force sensor head includes the cantilever,
a laser diode with suitable optics, and a position-sensitive detector. The preamplifier for this detector is
often located on the head. This minimizes the stray capacitance and hence the noise and maximizes the
bandwidth. The microscope shown in Figure 2.23 relies on external vibration damping. Common means
are an air-cushioned table or a concrete platform suspended by bungee cords.
If one is not satisfied with the scan range or the linearity of a microscope, then there is the possibility
to replace the base by a commercial linearized scanning table. The geometry is the same; all that is needed
is to adapt the support and the approach system of the force-sensing head.
AFMs operating in air usually have their vibration damping system built-in. Figure 2.25 shows the
multimode ultrahigh vacuum SPM of Omicron (Omicron; Howald et al., 1993). The instrument is based
on the optical lever principle (see Figure 2.24). The light beam coming from the laser diode is directed
by a mirror mounted on a piezomotor with two tilt directions to the cantilever. The reflected light is