
104 Measurement and Data Analysis for Engineering and Science
indicators of the variables sensed. Both the static and the dynamic response
characteristics of linear measurement systems are examined. First-order and
second-order systems are considered in detail, including how their output
can lag in time the changes that occur in the experiment’s environment.
With this information, approaches to data acquisition and signal processing,
which are the subjects of subsequent chapters, then can be considered.
4.2 Static Response Characterization
Measurement systems and their instruments are used in experiments to
obtain measurand values that usually are either steady or varying in time.
For both situations, errors arise in the measurand values simply because
the instruments are not perfect; their outputs do not precisely follow their
inputs. These errors can be quantified through the process of calibration.
In a calibration, a known input value (called the standard) is applied
to the system and then its output is measured. Calibrations can either be
static (not a function of time) or dynamic (both the magnitude and the
frequency of the input signal can be a function of time). Calibrations can be
performed in either sequential or random steps. In a sequential calibration
the input is increased systematically and then decreased. Usually this is
done by starting at the lowest input value and calibrating at every other
input value up to the highest input value. Then the calibration is continued
back down to the lowest input value by covering the alternate input values
that were skipped during the upscale calibration. This helps to identify
any unanticipated variations that could be present during calibration. In a
random calibration, the input is changed from one value to another in no
particular order.
From a calibration experiment, a calibration curve is established. A
generic static calibration curve is shown in Figure 4.1. This curve has several
characteristics. The static sensitivity refers to the slope of the calibration
curve at a particular input value, x
1
. This is denoted by K, where K =
K(x
1
) = (dy/dx)
x=x
1
. Unless the curve is linear, K will not be a constant.
More generally, sensitivity refers to the smallest change in a quantity that
an instrument can detect, which can be determined knowing the value of K
and the smallest indicated output of the instrument. There are two ranges
of the calibration, the input range, x
max
− x
min
, and the output range,
y
max
− y
min
.
Calibration accuracy refers to how close the measured value of a calibra-
tion is to the true value. Typically, this is quantified through the absolute
error, e
abs
, where
e
abs
= |true value − indicated value|. (4.1)