
Calibration and Response 125
These equations show that the system response will contain both magnitude
and phase errors. The magnitude and phase responses for different values of
ζ are shown in Figures 4.8 and 4.9, respectively. Note that the magnitude
ratio is a function of frequency, ω, for the sinusoidal-input forcing case,
whereas it is a function of time, t, for the step-input forcing case.
First examine the magnitude response shown in Figure 4.8. For low
values of ζ, approximately 0.6 or less, and ω/ω
n
≤ 1, the magnitude ra-
tio exceeds unity. The maximum magnitude ratio occurs at the value of
ω/ω
n
=
p
1 − 2ζ
2
. For ω/ω
n
≥∼ 1.5, the magnitude ratio is less than unity
and decreases with increasing values of ω/ω
n
.
Typically, magnitude attenuation is given in units of dB/decade or
dB/octave. A decade is defined as a 10-fold increase in frequency (any
10:1 frequency range). An octave is defined as a doubling in frequency (any
2:1 frequency range). For example, using the information in Figure 4.8, there
would be an attenuation of approximately −8 dB/octave [= 20log(0.2) −
20log(0.5)] in the frequency range 1 ≤ ω/ω
n
≤ 2 when ζ = 1.
Now examine the phase response shown in Figure 4.9. As ω/ω
n
increases,
the phase angle becomes more negative. That is, the output signal begins
to lag the input signal in time, with this lag time increasing with ω/ω
n
. For
values of ω/ω
n
< 1, this lag is greater for greater values of ζ. At ω/ω
n
= 1,
all second-order systems having any value of ζ have a phase lag of −90
◦
or
1/4 of a cycle. For ω/ω
n
> 1, the increase in lag is less for systems with
greater values of ζ.
4.7 Higher-Order System Dynamic Response
As seen in this chapter, the responses of linear, first, and second-order sys-
tems to simple step and sinusoidal inputs are rather complex. Most ex-
periments involve more than one instrument. Thus, the responses of most
experimental measurement systems will be even more complex than the
simple cases examined here.
When each instrument in a measurement system is linear, as described
in Chapter 2, the total measurement system response can be calculated
easily. For the overall system, [a] the static sensitivity is the product of all
of the static sensitivities, [b] the magnitude ratio is the product of all of the
magnitude ratios, and [c] the phase shift is the sum of all of the phase shifts.
In the end, the most appropriate way to determine the dynamic response
characteristics of a measurement system is through dynamic calibration.
This can be accomplished by subjecting the system to a range of either
step or sinusoidal inputs of amplitudes and frequencies that span the entire