n
-Propyl ether 65 0 1:37 2:06 1:15
40 0:43 0:29 0:18
n
-Propyl formate 50 0 0:46 1:07 2:10
40 0:17 0:18 0:27
Pyridine 128 0 8:00 4:15 –
40 – 1:26 –
Styrene 85 0 20:00 25:00 28:00
40 2:06 2:26 2:39
1,1,1,2-Tetrachloroethane 100 0 8:25 7:48 10:00
40 1:16 1:27 1:37
1,1,2,2-Tetrachloroethane 100 0 32:00 14:00 50:00
40 4:09 2:37 7:51
Tetrachloroethylene 70 0 3:00 5:45 2:10
40 0:41 1:06 0:33
Tetrahydrofuran 40 0 1:05 1:05 1:45
40 0:23 0:23 0:30
Toluene 110 0 2:30 4:05 4:30
40 0:38 0:47 0:53
1,1,1-Trichloroethane 105 0 0:31 1:10 0:47
40 0:15 0:23 0:20
1,1,2-Trichloroethane 85 0 5:13 3:43 15:00
40 0:40 0:45 2:30
Trichloroethylene 70 0 1:17 2:02 1:25
40 0:23 0:28 0:28
Trichlorotrifluoroethane (Freon 113) 80 0 0:15 0:30 0:16
40 0:13 0:14 0:13
Triethylamine 70 0 – 2:49 –
40 –––
Vinyl acetate 50 0 0:34 0:43 1:55
40 0:15 0:17 0:30
Vinylidene chloride 40 0 0:20 0:25 0:22
40 0:13 0:14 0:14
m
-Xylene 111 0 2:39 12:03 8:31
40 0:39 1:43 1:17
o
-Xylene 116 0 3:29 15:07 8:40
40 0:48 1:58 1:45
p
-Xylene 116 0 2:46 12:25 8:23
40 0:39 1:42 1:19
B 3% Diisodecyl phthalate on Chromosorb W, AW, 60/80 mesh
G 10% SP – 2100 on Supelcoport, 60/80 mesh
T 10% 1,2,3-Tris(2-cyanoethoxy) propane on Chromosorb P, AW, 60/80 mesh
(1)
Century organic vapour analysers are factory calibrated to measure ‘total organic’ vapours according to a standard
(methane). Since different organic vapours interact with the flame ionization detector (FlD) to varying extents, it is vital
that the instrument user be aware of the magnitude of the variation in order to obtain the most accurate data. Each user
must determine relative responses for the individual instrument.
(2)
For chromatographic work, the OVA can be used with a variety of column lengths and packing materials. For highest
accuracy, temperature control for the column is mandatory. This is accomplished using the portable isothermal pack (PlP)
kit which is supplied with three 8 in (203 mm) columns packed with B, G and T materials respectively. Isothermal control
is accomplished non-electrically using an ice-water mixture for 0°C and a seeded eutectic mixture for 40°C. The data
listed are for comparison purposes only since retention time for a compound can vary according to the condition of the
column packing material, packing procedure and chemical interaction among the components of a vapour mix.
A blank in the table indicates that no data are available for the analysis.
TOXIC PARTICULATES 333
Table 10.16 Cont’d
Column packing
retention time
(min:secs)
————————————————
Compound B-8 G-8 T-8
Relative
response
(1)
(%)
Column
temperature
(2)
(°C)