3.3 Overview of the instrument 141
In addition to the five main stages listed above, a TEM contains a number of
subsystems that are vital for proper operation, but are of secondary importance to
the operator during routine use of the microscope.
†
r
The high voltage system. This system is usually located in a separate unit and is connected
to the electron gun by a thick cable. For acceleration voltages of 400 kV and higher, this
type of cable connection is no longer possible and the accelerator is placed just behind
the microscope column or on top of the column.
r
The vacuum system. Electrons need a good vacuum (in the range of 10
−7
–10
−10
Pa (or
10
−5
–10
−8
torr, where 1 torr = 133.322 Pa) to travel the typically 1–2 m length of the
column, hence every TEM is equipped with a high vacuum system. The system generally
consist of one or more roughing pumps, possibly with a vacuum buffer tank, one or more
oil diffusion pumps and/or ion getter pumps and/or turbomolecular pumps, and a num-
ber of independent vacuum circuits with manual or automatic (pneumatic or electronic)
valves. Some microscopes are modified for ultra-high vacuum conditions, which means
that the partial pressures of all gases in the column are better than 10
−8
torr. In older
microscopes, such as the Siemens Elmiskope 102, the operator had to be familiar with
the pumping diagrams and cycles and often manually open or close valves. Nowadays,
this task is fully automated and interlocks on the high voltage system and the filament
emission system will prevent the operator from using the machine if the vacuum level is
inadequate.
The operator should always take note of the proper vacuum level at which the micro-
scope must be operated. This should be part of the prework checklist. The vacuum is
usually best in the electron gun and slightly worse at the specimen stage. Some micro-
scopes allow observations at near ambient pressures in a variety of gas mixtures through
the use of an environmental specimen cell; in those cases the sample region is separated
from the rest of the microscope column by differential pumping apertures.
In older microscopes, such as the JEOL 120CX, the vacuum region of the microscope is
separated from the rest by about two-dozen O-rings. In modern instruments the electrons
travel through so-called liner tubes, long cylindrical sections with a small inner diameter
(a few millimeters), which extend across several lenses and reduce the total number of
vacuum seals needed.
r
The cooling system. Large electric currents can generate a lot of heat (Ohm’s law), hence
the lens coils, which can carry several amperes of current, must be water-cooled. This
system is fully automatic and the operator should never have to intervene at any point.
The cooling system is interlocked with other systems, so that the microscope cannot be
used if insufficient cooling water is available.
r
Radiation shields. All electrons emitted by the filament must eventually find their way
back into the electrical circuit, and must get rid of their kinetic energy. All surfaces
†
A TEM operator should be aware of the meaning of the warning signs and indicator lights should any of the
subsystems described in this section malfunction or cause problems. We refer to the microscope manual for a
detailed description of all subsystems.