
2.3 Liquid Chromatography Techniques and Applications in Proteome Analysis 177
Tab. 2.7: Relationship between column i.d. and sensitivity.
Scale Column i.d.
(mm)
Typical flow rate
(mL/min)
Theoretical gain in
sensitivity
Analytical 4.6 1,000 1
Narrow 2.1 200 5
Micro 1 40 21
Capillary 0.3 4 237
Nano 0.075 0.2 3,322
If the sample of interest is rather simple, e.g. a tryptic digest of a
gel spot, ideally containing a single or a few proteins, successful pro-
tein identification with a very high sequence coverage can be routi-
nely achieved with a single dimension of RPC separation prior to MS.
However, as a sample becomes more complex the required resolu-
tion of the RPC separation prior to tandem mass spectrometry
sequence analysis also needs to increase. In this instance, a single
RPC separation prior to MS analysis may still generate a significant
number of protein identifications, but the sequence coverage may
drop considerably (for instance an immunoaffinity pull down experi-
ment, followed by digestion of the 1DE gel bands – see introduction
and figure 4). The question becomes, does the MS have sufficient
time to analyze the peptides (i.e. perform sequence analysis) as they
co-elute from the column? One possibility to address the co-elution is
to employ extra high-resolution RPC with very long columns, where
the system pressure approaches 10,000 psi (700 bar; 6,895 kPa). In
one such publication, Shen et al. (2001) demonstrated that a single
high-resolution RPC separation prior to tandem mass spectrometry
with an FT-ICR instrument, yielded the identification of over 1,000
yeast proteins. Instead of following the approach with very long col-
umns, which has never got a wide acceptance, one could use shorter
columns packed with very small particles (1–2 mm) instead. Such a
strategy, marketed as UPLC, is more promising and is currently gain-
ing some popularity. Both, columns and dedicated instrumentation,
are commercially available.
If the sample complexity exceeds a certain level, such as with a
sample deriving from of a whole cell lysate digest, the ability to effi-
ciently separate the mixture into manageable fractions with one
dimension only becomes effectively impossible. This is critical as the
ability of the MS to analyze the co-eluting peptides will be seriously
impaired; not only will the sequence coverage of identified proteins
be reduced, but insufficient peptides will be detected and analyzed to
deliver unambiguous protein identifications, significantly reducing
Ideally, such a simple sample
could be analyzed with nano-
spray alone.
A high-speed MS analyzer is
beneficial.
Shen Y, Zhao R, Belov ME.
Anal Chem 73 (2001) 1766–
1775.
Alternatively, the resolution of
an RPC step can be enhanced
by simply increasing the length
of the RPC gradient within
certain limits (0.25% acetoni-
trile/min increase should be the
flattest gradient).
This is particularly important
for the analysis of peptides
deriving from low-abundant
proteins.