
The effect of a spatially variable radiation flux falling on a pyroelectric plate is
to produce a corresponding charge distribution. In imaging devices this charge
distribution is detected by a process giving rise to a current which is amplified
and electronically processed to produce a television picture corresponding to the
infrared radiation from the original scene. A widely used device is the infrared
vidicon shown schematically in Fig. 7.11. It is used in many systems, both civil
and military. A UK-manufactured camera is used for locating survivors buried
in collapsed building rubble.
The radiation from the scene is chopped and focused, using germanium optics,
onto an infrared absorbing electrode carried on the front face of the pyroelectric
plate. The voltages developed across the plate are typically a few millivolts. The
metal grid, positioned close to the detector, collects most of the electrons as it is
scanned. Some electrons pass through the grid; the number depends on the local
potential at the surface of the plate. The charge deposited on the plate produces a
signal at the front electrode which is processed into the television picture.
In the particular mode of operation outlined, when the radiation is interrupted
by the chopper, the detector cools, giving negative signals and causing the local
surface potential to go negative with respect to the cathode. In order for the
detector to be properly addressed by the electron beam it is biased to a potential
of approximately þ300 mV. This is achieved by the deposition of charge from
gas molecules deliberately introduced into the tube and ionized by the electron
beam. Circuitry in synchronism with the chopper reverses the sense of the
negative output signal current when the chopper is closed and adds this to the
positive signal obtained when the chopper is open. This both removes any flicker
in the picture due to detector or bias non-uniformity and increases the sensitivity
for imaging moving objects.
The charge on the surface of the plate must be removed before the electron-
beam scan is repeated. This will occur automatically if the bias is high enough to
ensure that sufficient electrons can reach the plate to neutralise the pyroelectric
APPLICATIONS 427
Fig. 7.11 Pyroelectric vidicon.