microscope is over-focused and the needle lowered and brought into focus.
This process is repeated until the needle’s tip is positioned above the FIB
cuts. The sample is then rotated to align the lamella parallel to the needle’s
shaft. To illustrate the lift-out steps SE images, from when the technique has
been performed within an FIB/SEM system, are used (this alternative
approach is discussed at the end of this section) as these are more informative
than the equivalent optical images. The needle is further lowered and
repeatedly swept through the FIB cuts and pushed against the lamella until
the latter sticks to it through electrostatic forces (Figures 8.3 (d) and (e)).
Once the lamella is attached to the needle, the needle is raised away from the
sample and the sample is replaced with a carbon coated TEM grid. The
needle and lamella are then lowered (again by focusing on the needle tip,
over-focusing, and lowering the needle into focus) and then repeatedly swept
across the surface of the carbon membrane until the lamella sticks to the
carbon. Generally, the adhesion between the lamella and carbon membrane
is greater than that between the needle and lamella. If necessary, the needle is
then used to push the lamella flat onto the carbon (Figure 8.3 (f)). The
adhesion between the lamella and carbon is sufficiently strong that only
rarely does a TEM lamella fall off during transportation to or mounting in
the TEM. As well as continuous carbon membranes, holey and lacy carbon
films can be used [50, 51]. However, adhesion of the lamella to these supports
is not as good (due to the smaller contact area) and typically 1 in 30 samples
may be lost while transferring the grids to the TEM.
The success rate for the lift-out step ranges from 60% to 90%. Failure can
occur when the lamella is being lifted out from the FIB cuts or as it is being
placed onto the supporting membrane. During the lifting out step, the
lamella may ‘‘jump’’ up the shaft of the needle, away from the tip, such that it
cannot be brushed against the carbon membrane. Several approaches have
been used to reposition the lamella at the needle’s tip. These include (i) using
a second micromanipulator, (ii) brushing the needle against the rim of the
TEM grid, and (iii) tapping the micromanipulator so that the lamella jumps
off the needle and onto the support membrane. Of these three methods,
tapping the micromanipulator is probably the quickest and easiest one to do.
The lift-out can also fail if the carbon membrane is accidentally punctured
and ripped by the tip of the needle as the lamella is being placed onto it
(Figure 8.4 (a)). This can result in (i) the carbon support membrane rolling
around and encasing the lamella, (ii) the lamella on the ripped carbon
membrane being at a large angle to the electron beam in the TEM such that it
cannot be tilted onto a crystal axis (Figure 8.4(b)), and (iii) the lamella falling
through the resulting hole and being lost. To prevent these problems TEM
Focused ion beam systems224