portion of the in-lens detector, shown on the right-hand side of Figure 6.5.
Those BSEs with energy higher than the grid voltage navigate across the field
and are directly detected by the EsB detector. The grid voltage is not only
used to reject secondary electrons but it is also an energy filtering element
that can be used to create an energy window between the primary and grid
voltages. Using this type of BSE
I
detection system it is possible to collect a
low voltage high spatial resolution z-contrast image at short working distance
with a 25 eV energy image window. Isolating the BSE
I
signal at low voltage
coupled with energy filtering to concentrate on the elastic peak is a benefit in
3D reconstruction.
High scattering angle BSE
I
images will also display minimum topography in
addition to strong material contrast, as illustrated through the image pair in
Figure 6.6. The sample is an uncoated geological material. The images were
acquired at 750 V with a horizontal width of 7.7 m m. Both images in Figure 6.6
were acquired simultaneously during a single scan and captured using two
detectors. The side chamber mounted SE
2
detector was used to collect the
image shown in (a) and the energy selective backscatter (EsB) detector was
used to collect the image in (b). The SE
2
detector image shows strong topo-
graphic contrast while the EsB image displays minimum topography and
strong z-contrast, in a dramatic example of the different information content
extracted by the two detectors. Other contrast mechanisms such as Schottky
barrier contrast, voltage contrast and charging contrast are also suppressed
from an EsB image. Enhanced z-contrast with minimum topography is a
benefit in FIB-SEM 3D reconstruction to separate different materials and
phases throughout the 3D volume during data reduction. Similarly, the elim-
ination of edge contrast, charging contrast, and other contrast mechanisms
deleterious to FIB-SEM nanotomography is also a significant benefit. The
ability to extract this information at low voltages of 1 kV and less means that
depth resolution is also optimized. Moreover, this type of in-lens BSE detector
requires no insertion, alignment, or adjustment. Thus, in many respects the in-
lens EsB is an ideal detector for the purposes of FIB-SEM 3D nanotomo-
graphy. It is expected that future work will continue to highlight the benefits of
FIB-EsB nanotomography in both materials science and emerging biological
applications.
Electrostatic–electromagnetic objective lens
The Gemini column derives it name from the ‘‘twin’’ element configuration of
the objective lens. With attention to Figure 6.3(b), the upper portion of the
objective lens is an electromagnetic element while the lower portion of the
objective lens is an electrostatic element. Collectively, these elements combine
Focused ion beam systems160